UL120

Hybrid Sensor Input Data Logger
 

Product Outline

UL120 is a logger to record various types of physical data.

UL120 is a logger to record various types of physical data including weather information such as rainfall, snow, wind direction and wind velocity. Various sensors can establish a direct connection without using a dedicated amplifier. Operation can be performed directly or setting and data transfer can be done via communication by RS-232C or cell-phone. Since collected data can be recorded in a CF (compact flash) card in CSV format, compatibility with PC is outstanding.

製品イメージ
 

Recording data sample icon 

 
 

Specifications

INPUT/OUTPUT SECTION (analog input)
Number of channels single-end 16ch, or differential input 8ch (can be mixed)
*1-2ch, 3-4ch, ~ 15-16ch setting at differential input
Input range ±5mV, ±50mV, ±500mV, ±5V
Input type DC voltage, thermocouple (K, E, J, T, R) pt100
Bridge (6-wire, 4-wire full, 3-wire, 2-wire, AC half)
Input resistance 10MΩ
Display data Voltage measurement…5-digit display (max. 99999), free decimal point position
Temperature measurement…5-digit display (max. 99999), display resolution 0.1℃
Temperature measurement range
  • Thermocouple…K -270~1370℃ E -270~750℃ J -210~990℃ T -270~400℃ R -50~1760℃
    Pt100 …-200~650℃
Accuracy
  • Voltage input (5V, 500mV, 50mV range) :
  • ±0.08%FS(short mode)
    ±0.033%FS(normal mode)
  •  
  • Voltage input (5mV range) :
  • ±0.3%FS(short mode)
    ±0.1%FS(normal mode)
  •  
  • Thermocouple input (K, E, J, T) :
  • within ±0.05% rdg, ±1.0℃
  •  
  • Thermocouple input (R) :
  • within ±0.1% rdg, ± 3.0℃
  •  
  • Pt100 input :
  • within ±0.1% rdg, ±1.0℃ (short mode)
    within ±0.1% rdg, ±0.5℃ (normal mode)
Simultaneity
  • Short mode :
  • 10 msec or less
  •  
  • Normal mode :
  • 1 sec or less
 
INPUT/OUTPUT SECTION (Pulse input)
Number of channels For 8-bit counter channels, or two 16-bit counter channels (mixable)
For 16-bit counters, use 1-2ch and 3-4ch in pairs.
Input type TTL、OPEN C、CONTACT、AC LOW LEVEL
Measurement frequency range ■When used as frequency counter (duty 50%)
- TTL, OPEN C, CONTACT
High-frequency: Instantaneous frequency during sampling
Low-frequency: Value obtained by dividing integrated value by interval time
  • - 16 bit high-frequency:
  • 100 Hz - 250 kHz (by 100 Hz)
  •  
  • - 16 bit low-frequency:
  • 1 Hz - 65 kHz (by 0.1 Hz, 1 sec sampling -)
  •  
  • - 8 bit high-frequency:
  • 100 Hz - 25.5 kHz (by 100 Hz)
  •  
  • - 8 bit low-frequency:
  • 1 Hz - 255 kHz (by 0.1 Hz, 1 sec sampling -)

- AC LOW LEVEL
Zero crossing comparator used, hysteresis, approx. 1.5 mV
  • ・20mVrms:
  • 1Hz - 100Hz
  •  
  • ・50mVrms:
  • 0.5Hz - 400Hz
  •  
  • ・150mVrms:
  • 0.3Hz - 1kHz

■When used as accumulation counter
  • - 16bit:
  • 65535 counts
  •  
  • - 8bit:
  • 255 counts
 
INPUT/OUTPUT SECTION (digital input)
Number of channels 8 ch
Input voltage TTL,OPEN C,CONTACT
 
INPUT/OUTPUT SECTION (event input)
Number of channels 1 ch
Detection method Falling edge (3.5V or more →1.5V or less)
 
INPUT/OUTPUT SECTION (sensor feeding power output)
Number of channels +5V: 4 ch, +12V: 4 ch, Firm power (excluding SLEEP time)
Output voltage +5V fixed, +12V
Voltage accuracy +5V :+5% -10%
+12V: Dependent on UL120 supply voltage (external power, battery)
Output current 100mA/ch
 
INPUT/OUTPUT SECTION (exciter output)
Number of channels 4 ch, Output in synchronization with data acquisition
Output voltage Settable within +5V
Type of constant output on plus side and type of output on plus and minus sides in rectangular wave
Output current 30mA / ch (at 5V output)
Accuracy ±0.1%
 
INPUT/OUTPUT SECTION (analog output)
Number of channels 2 ch
Output voltage Output in ±5V range for any input channel (Output ON/OFF settable)
Accuracy ±0.1%
 
INPUT/OUTPUT SECTION (digital output)
Number of channels 8 ch
Output level TTL output
Pit assign CH1: Logging (H)
CH2: 1st logging started (H)
CH3: Power to this equipment is ON during logging in (H)
CH4: Data sampling (H)
CH5: Any input exceeds alarm high limit value (H)
CH6: Any input exceeds alarm Low limit value (H)
CH7: reserve
CH8: reserve
 
DATA RECORDING SECTION
Recording media PCMCIA ATA-FLASH card (PCMCIA R2.1 JEIDA Ver 4.2 standard compliance product)
*Compact Flash TM card usable with adapter
FAT file system DOS interface guideline Ver. 1.1 compliance
File type Text file in CSV format (usable in spreadsheet software such as MS-Excel)
Record contents Time data: Start time, finish time, finish state, etc.
Summary data: Logging time (recording: year, month, day, hour, min, sec), logging frequency
Logging data: Individual acquisition data
 
INTERFACE SECTION
Communication standard RS-232C compliance
Transmission rate 38400、19200、9600、4800 bps
Bit configuration Stop: 1/2 bit
Data length: 7 / 8 bit
Parity: NONE/ODD/EVEN
Connector D-sub 9pin
 
TIME
Display / record Year, month, day, hour, min, sec (Year: last 2 digits of western calendar)
Accuracy Month difference one minute (normal temperature)
 
POWER SECTION (Internal battery or external battery, whichever voltage value is higher automatically selected)
Internal battery
  • ・Battery used
  • Lead battery (2.3Ah), usable for approx. 2 months at 10 min. intervals (according to use conditions)
  •  
  • ・Charge mode
  • Charge by separately-sold battery charger (approx. 10 hours) Internal trickle charge when AC adapter is used (approx. 48 hours)
External power dedicated AC adapter (15V DC) or external battery, Input voltage..12 - 16V
Backup battery Setting value retained by lithium battery (Retention period: 60 months)
 
DISPLAY SECTION
Display unit LCD module, Time / Setting value / Input data etc.
Operation display LED (orange) x 1: Flashing during logging
 
OPERATION SECTION
Membrane key Operation such as setting value input, card formatting
Switch ・ POWER (Power switch): Main power supply of UL120 ON/OFF
・ START (Start switch): Logging starts when switch is ON.
 
FUNCTIONS
Data recording method
  • ・Interval mode
  • Logging every set interval time.
    Short: 10msec - 990 msec (by 10 msec) Normal: 1 sec - 24 hours (by 1sec)
  •  
  • ・Event mode
  • Logging when event signal is input
Data record value
  • ・Instantaneous value
  • Instantaneous value at the time of recording is recorded.
  •  
  • ・Statistical value
  • Only average, maximum, and minimum values of data in the section set from the time of recording are recorded.
Specified time (frequency) complete function Logging is completed when the set time (frequency) is reached after logging.
Scale setting
  • ・Zero setting
  • Any voltage value can be set as 0 display.
  •  
  • ・Span setting
  • Any voltage value can be set as any display value (physical value).
Analog input calculation function Function to calculate (addition, subtraction) set values between arbitrarily-set analog 2 channels and record as one data
 
GENERAL SPECIFICATIONS
Operating conditions Accuracy assurance temperature range: 0℃ - 40℃ (Humidity: 90% RH or less, non-condensing)
Working temperature range: -10℃ - 50℃ (Humidity; 90% RH or less, non-condensing)
External dimensions 220 (W) x 60 (H ) x 195 (D) mm (protruding parts not included)
Weight Approx. 2.2 kg (battery included)
 
ACCESSORIES
AC adapter 1
Mini-driver 1
Seal-type lead battery 1
Operation Manual 1
 
 

Optional Accessories

Model Description
AP1516 AC adapter for UL120 (OUT: 15V, 1.6A)
P1205 Battery charger for UL120 (10 hours chargeable)
BP1223 Battery pack for UL120 (seal type lead battery 12V, 2.3Ah)
CA120-232X D-sub9p-D-Sub9p crossing cable 1.5m
CF 128M CF card 128MB
CF 256M CF card 256MB
CF 512M CF card 512MB
CF-ADP CF card adapter
 
 

Download

Product CatalogueOperation ManualsExternal dimentionsSupport ToolsSoftware
UL120
Rev.1.05(1.36MB)
 
  • Download an operation manual, product appearance diagram, support tool or software after user registration.